| Specifications for specific thermocouple types | ||||||
Type |
K |
N |
R |
S |
B |
J |
Measuring range |
-100°C
1370°C -148°F 2498°F |
-100°C
1300°C -148°F 2372°F |
-40°C
1760°C -40°F 3200°F |
-40°C
1760°C -40°F 3200°F |
55°C
1815°C 131°F 3300°F |
0°C
800°C 32°F 1472°F |
Accuracy |
±
0.3°C / 0.5°F |
±
0.4°C / 0.7°F |
±1.0°C at 200°C ±1.8°F at 392°F ±0.7°C at 1000°C ±1.3°F at 1832°F |
±1.0°C at 200°C ±1.8°F at 392°F ±0.8°C at 1000°C ±1.4°F at 1832°F |
±1.0°C
at 1500°C ±1.8°F at 2732°F ±3°C at 400°C ±5.4°F at 752°F |
±0.2°C
/± 0.4°F |
Resolution |
0.1°C
/ 0.2°F |
0.1°C
/ 0.2°F |
0.1°C
/ 0.2°F |
0.1°C
/ 0.2°F |
0.1°C
/ 0.2°F |
0.1°C
/ 0.2°F |
| Specifications for specific logger types | ||||
| Normal operating temperature | High operating temperature | |||
| Model number* | TP2086 | TP2016 | TP2186 | TP2116 |
| No. of channels | 8 | 10 | 8 | 10 |
| Memory (data points) | 130,000 | 130,000 | 130,000 | 130,000 |
| Max. operating temp. | 70°C / 158°F | 70°C / 158°F | 110°C / 230°F | 110°C / 230°F |
| Battery type | NiMH rechargeable | NiMH rechargeable | VHT Lithium | VHT Lithium |
| Max. battery life between charges** | 340 hours | 340 hours | 250 hours | 250 hours |
| Sampling interval: | ||||
|
No
telemetry |
0.1 sec 50 mins | 0.1 sec 50 mins | 0.1 sec 50 mins | 0.1 sec 50 mins |
|
Hardwired
telemetry |
1 sec 50 mins | 1 sec 50 mins | 1 sec 50 mins | 1 sec 50 mins |
|
RF
telemetry |
n/a | 2 secs 50 mins | n/a | 3 secs 50 mins |
| Max. operating pressure at 20°C / 68°F | 20 bar / 290 psi | 20 bar / 290 psi | 20 bar / 290 psi | 20 bar / 290 psi |
| Min. operating pressure at 20°C / 68°F | 10-5millibar | 10-5millibar | 10-5millibar | 10-5millibar |
| *
Model number shown is for type K thermocouples only, other types have different
model numbers ** Max. battery life depends on sampling interval, operating temperature and use of RF telemetry |
||||
| Case size (all logger types) | |
| Length | 200 mm / 7.87" |
| Width | 98 mm / 3.86" |
| Height | 20 mm / 0.79" |
| General operating specifications | |
| Logger start by: | Manual (button), Temperature (rising or falling), or Time |
| Programmable events: | Up to 8 sampling interval changes possible per run, trigger by rising or falling temperature after first event |
| Hot data protection: | By non-volatile memory, and software warning if reset attempted before download |
| Communication: | Via USB or RS 232 for lap top computers |
| Logger reset: | Possible after download by start button (uses same run parameters) |